Description
Appliance:
Temperature Humidity Climatic Test Chamber Price is applied to test the materials in heat-resistance, cold-resistance, dry-resistance, moisture-resistance. It is simple to operate and program easy to edit. It can show the set values and operative time.
Temperature Humidity Climatic Test Chamber Price applied to test the prodcuts quality, such as electronic, plastic products, electrical appliances, instruments, food, vehicles, metals, chemicals, building materials, aerospace, medical care and so on.
Standards:
The performance index meets the requirements of GB5170, 2, 3, 5, 6-95 verification methods for basic parameters of
environmental test equipment for electrical and electronic products low temperature, high temperature, constant humidity and alternating humidity heat test equipment
Basic environmental test procedures for electrical and electronic products test a: low temperature test method GB 2423.1-89 (IEC68-2-1)
Basic environmental test procedures for electrical and electronic products test B: high temperature test method GB 2423.2-89 (IEC68-2-2)
Basic environmental test procedures for electrical and electronic products test CA: constant damp heat test method GB/T2423.3-93 (IEC68-2-3)
Basic environmental test procedures for electrical and electronic products test Da: alternating damp heat test method GB/T2423.4-93 (IEC68-2-30)
ETC…
Refrigeration system:
Compressor: fully hermetic compressor imported from France
Refrigerant: Environmental refrigerant R404A
Condenser: fin type with heat dissipation motor
Evaporator: fin type multi section automatic load capacity adjustment
Other accessories: desiccant, refrigerant flow window, repair valve
Expansion system: refrigeration system with capacity control
Control system:
Imported from Korea, all Chinese LCD 7 “LCD touch panel, drawing face to talk input data, temperature and humidity can be programmed at the same time, curve display, set value / display value curve. It supports the direct export of temperature and humidity record data by U disk, and can display various alarms respectively. When the fault occurs, it can display the fault on the screen to eliminate the fault and misoperation. Multi group PID control function, precise monitoring function, and display on the screen in the form of data.
Controller specification:
1. Accuracy: temperature ± 0.1 ℃ + 1digit, humidity ± 1% RH + 1digit
2. Resolution: temperature ± 0.01, humidity 0.1% R.H
3. With upper and lower limit standby and alarm functions
4. Temperature and humidity input signal selection PT 100 Ω× 2 (dry bulb and wet bulb)
5. Temperature and humidity conversion output: 4 ~ 20mA
6. Six groups of P.I.D control parameter setting, P.I.D automatic calculation, dry and wet bulb automatic correction
Specifications:
Item Name | Temperature and humidity test chamber/machine | |
Capacity(L) | 80, 150, 225, 408, 500, 800, 1000L(optional, or custom-made) | |
Performance | Control Methods | Balanced temperature humidity control system (BTHC) PID intelligent adjust |
Temperature and Humidity Range | -70°C~150°C; 20%-98%R.H | |
Accuracy | ±1.0°C, ±2%RH / ±0.5°C, ±2%RH | |
Resolution | ±0.1°C, ±0.1%RH / ±1°C, ±3%RH | |
Heating Rate | ≤3°C/min | |
Cooling Rate | Average 1°C/min | |
Internal Material | SUS#304 stainless steel | |
External Material | SUS#304 stainless steel or Antirust processing cold-rolled steel sheet(Spray) | |
Insulation Material | Superfine wool + polyurethane foam | |
Regulator | Cooling Mode | Mechanical binary refrigeration way |
Compressor | Closed compressor (French Tecumseh or German Copeland ) | |
Refrigerant | R23/R404A | |
Condensing Way | Air-cooling and Water-cooling | |
Humidifier | Stainless steel humidifying tube | |
Airflow Way | Broadband-type forced air circulation | |
Controller | Operate Interfaces | Korea TEMI 880 LCD touch screen |
Memory capacity | 120 groups, each program 100 segment | |
Input | PT100 or T thermocouple | |
Communicate Interface | RS-232 | |
Subsidiary function | Upper and lower alarm; self diagnosis ; alarm display(fault cause); time sequence protection (Automatic turn on or turn off ) | |
Water supply | Water supply | Circulating water supply |
Water tank | Built-in hand type(30Kg capacity) | |
Water quality | resistance >500Ω.m | |
Standard Configuration | One observe window(Double hollow toughened glass);one test pattern D50mm(on the left); 2 layers sample frame; one floodlight(fluorescent lamp);one water feeding tank; | |
Safety device | Power leakage protection; Dry heating protection; Water protection Over-temperature protection switch; Temperature deviation alarm, Compressor Overpressure; overload; Water shortage; Phrase shortage, etc. | |
Power supply | 220VAC ±10%50/60Hz & 380VAC ±10%50/60Hz |
Reviews